The Food and Drug Administration on Wednesday, Nov. 25, posted a recall notice for Tandem Mobi insulin pumps and called it the most serious type of recall. Using one of the pumps without following the company's instructions may cause serious injury or death, the agency says. Tandem Diabetes Care has reported 1,248 serious injuries and no deaths tied to the problem as of Oct. 28.

The recall covers every Tandem Mobi pump with interoperable technology, the version that can link to a continuous glucose monitor and adjust insulin automatically. The FDA's recall database puts the number at 80,938 pumps, including 77,559 in the United States, sold nationwide and in Canada.

What goes wrong

A man wears an insulin pump clipped to his belt, with clear tubing running to an infusion site on his abdomen and a glucose sensor on his upper arm
An insulin pump with tubing to an infusion site and a glucose sensor on the arm, photographed in 2022. The pump shown is not a Tandem model.Medtd1 / Wikimedia Commons / CC BY-SA 4.0

Some pumps sound an occlusion alarm, the warning that insulin is blocked, when nothing is blocking it. Tandem traced the false alarms to extra friction in the drive module, which can sap motor power until the motor runs slower than the pump's blockage threshold.

When the alarm sounds, the pump stops delivering insulin. It stays stopped until the wearer clears the alarm and restarts delivery. Repeated false alarms that go unanswered can interrupt insulin long enough to push blood sugar up, and Tandem says severe high blood sugar may need medical care or a hospital stay.

What Mobi users should do

An open gray travel case on a kitchen counter holding insulin pens, a vial, wrapped syringes and a glucose meter
Image: Edit USA

Tandem emailed customers on Sept. 23. What to do depends on the pump's serial number.

If your serial number is on the FDA's list of pumps subject to replacement, a spreadsheet download, request a return and replacement through Tandem's recall page or by calling Tandem Customer Technical Support at 877-801-6901. Keep following the steps below until the new pump arrives.

All other Mobi users can keep using their pumps if they:

  • Respond to every occlusion alarm: acknowledge it, check the cartridge, tubing and infusion site for damage or a blockage, fix any problem and resume insulin delivery.
  • Turn on push notifications in the Tandem Mobi app, so an alarm also shows up on the phone. Tandem has a support article on how.
  • Check blood sugar regularly for unexpected highs or lows.
  • Keep a backup way to take insulin, as their doctor directed.

Anyone getting an unusual number of occlusion alarms, or alarms that keep interrupting treatment after troubleshooting, should contact Tandem at 877-801-6901 or techsupport@tandemdiabetes.com.